Strategic Communications and Marketing News Bureau

Femtogram-level chemical measurements now possible, U. of I. team reports

William King, professor of mechanical science and engineering, left; Rohit Bhargava, professor of bioengineering; and Keunhan Park, postdoctoral research associate, have demonstrated a method for simultaneous structural and chemical characterization of samples at the femtogram level (a femtogram is one quadrillionth of a gram) and below.

William King, professor of mechanical science and engineering, left; Rohit Bhargava, professor of bioengineering; and Keunhan Park, postdoctoral research associate, have demonstrated a method for simultaneous structural and chemical characterization of samples at the femtogram level (a femtogram is one quadrillionth of a gram) and below.

CHAMPAIGN, Ill. – Finding a simple and convenient technique that combines nanoscale structural measurements and chemical identification has been an elusive goal. With current analytical instruments, spatial resolution is too low, signal-to-noise ratio too poor, sample preparation too complex or sample size too large to be of good service.

Now, researchers at the University of Illinois have demonstrated a method for simultaneous structural and chemical characterization of samples at the femtogram level (a femtogram is one quadrillionth of a gram) and below.

The measurement technique combines the extraordinary resolution of atomic force microscopy and the excellent chemical identification of infrared spectroscopy.

“We demonstrated that imaging, extraction and chemical analysis of femtogram samples can be achieved using a heated cantilever probe in an atomic force microscope,” said William P. King, a Kritzer Faculty Scholar and professor of mechanical science and engineering.

King and colleagues describe the technique in a paper accepted for publication in the journal Analytical Chemistry, and posted on its Web site.

The new technique hinges upon a special silicon cantilever probe with an integrated heater-thermometer. The cantilever tip temperature can be precisely controlled over a temperature range of 25 to 1,000 degrees Celsius.

Using the cantilever probe, researchers can selectively image and extract a very small sample of the material to be analyzed. The mass of the sample can be determined by a cantilever resonance technique.

To analyze the sample, the heater temperature is raised to slightly above the melting point of the sample material. The material is then analyzed by complementary Raman or Fourier transform infrared spectroscopic imaging, which provides a molecular characterization of samples down to femtogram level in minutes.

“Fourier transform infrared and Raman spectroscopic imaging have become commonplace in the last five to ten years,” said Rohit Bhargava, a professor of bioengineering. “Our method combines atomic force microscopy with spectroscopic imaging to provide data that can be rapidly used for spectral analyses for exceptionally small sample sizes.”

To clean the tip for reuse, the tip is heated to well above the decomposition temperature of the sample – a technique similar to that used in self-cleaning ovens.

“Since the tip can be heated to 1,000 degrees Celsius, most organic materials can be readily vaporized and removed in this manner,” King said.

As a demonstration of the technique, the researchers scanned a piece of paraffin with their probe, and removed a sample for analysis. They then used Raman and Fourier transform infrared spectroscopy to chemically analyze the sample. After analysis, the paraffin was removed by thermal decomposition, allowing reuse of the probe.

“We anticipate this approach will help bridge the gap between nanoscale structural analysis and conventional molecular spectroscopy,” King said, “and in a manner widely useful to most analytical laboratories.”

With King and Bhargava, co-authors of the paper are postdoctoral researcher and lead author Keunhan Park and postdoctoral research associate Jung Chul Lee. All four researchers are affiliated with the university’s Beckman Institute.

The work was funded by the National Science Foundation through the Center for Nanoscale Chemical-Electrical-Mechanical Manufacturing Systems, and by the U. of I.

Editor’s note: To reach William King, call 217-244-3864; e-mail: wpk@illinois.edu.



This article was imported from a previous version of the News Bureau website. Please email news@illinois.edu to report missing photos and/or photo credits.

Read Next

Earth and Environmental Sciences Researcher portrait standing in front of graphics from study

Researchers advance first-of-its-kind AI tool for translating life-saving weather warnings across the US

A new study led by Joseph Trujillo-Falcón documents how artificial intelligence is used to translate life-saving weather forecasts and alerts into non-English languages.

Arts Photo of a group of violinists from Apollo's Fire standing and playing onstage.

Krannert Center announces performers for 2026-27 season

Krannert Center for the Performing Arts’ 2026-27 season features a variety of performers, including jazz and classical music, theater and dance.

Life Sciences Research News Veterinary Medicine Photo of researchers in the laboratory. They are standing in front of dozens of fish tanks.

One simple trick makes zebrafish a better model for microbiome research

A new advance in animal husbandry involving a popular aquarium fish should speed the pace of microbiome research, scientists report.

Strategic Communications and Marketing News Bureau

507 E. Green St
MC-426
Champaign, IL 61820

Email: stratcom@illinois.edu

Phone (217) 333-5010